1
|
PROBLEM: |
|
| |
DIVERTER
LEAKING AT SPRAY ARM |
|
| |
CAUSE: |
|
|
A) Diverter
spray arm (old style) is damaged. |
|
B) Spray
arm o-ring is damaged or installed improperly. |
|
|
SOLUTION: |
|
|
A) Replace
diverter arm with new spray arm replacement kit #
08093. NOTE: On
old style diverter spray arm, the spray arm o-ring was placed
on groove at the end of the spray arm. When installing new
spray arm kit, place washer into bottom of hole of nozzle pivot.
Then place the o-ring on top of washer. Insert new spray arm
into the pivot, pressing in slightly. While holding spray arm
in place, tighten set screw. |
|
B) Re-Install
o-ring as described above in step (A.) If leakage persists,
replace using spray arm kit # 08093 |
|
|
| 2 |
PROBLEM: |
|
|
DIVERTER
LEAKING AT NOZZLE PIVOT. |
|
|
CAUSE: |
|
|
A) One
or both pivot o-rings are worn. |
|
B) Diverter
seal washer worn or missing. |
|
C) Nozzle
pivot damaged or defective. |
|
|
SOLUTION: |
|
|
A) Replace
nozzle pivot assembly. |
|
B) Replace
pivot seals. Use diverter rebuild kit. |
|
C) Replace
nozzle pivot assembly. Use pivot replacement kit. |
|
3 |
PROBLEM: |
|
|
ERRATIC
SPRAY PATTERN FROM SPRAY ARM NOZZLE |
|
|
CAUSE: |
|
|
A) Debris
clogging nozzle screen. |
|
B) Damaged
or missing nozzle screen or spray disc. |
|
|
SOLUTION: |
|
|
A) Wash
nozzle screen under running water. |
|
B) Replace
using # 08846 spray arm repair kit. |
|
|
|
|
4 |
PROBLEM: |
|
|
DEOSEPTIC
SOLUTION NOT SIPHONING FROM BOTTLE DURING FLUSH CYCLE (Deoseptic
models only.) |
|
|
CAUSE: |
|
|
A) Adjusting
screw not open. |
|
B) Aspirator
tube screen clogged. |
|
C) Aspirator
tube not installed completely onto swivel & ball check
assembly. |
|
D) Aspirator
tube cracked. |
|
E) Rubber
ball check stuck inside swivel & ball check assembly. |
|
| |
SOLUTION: |
|
| |
A) Open
adjusting screw until adequate amount of solution is siphoned
from bottle during flush cycle. |
| |
B) Remove
solution bottle. Remove aspirator tube from swivel & ball check
assembly and wash screen under running water. |
| |
C) Remove
solution bottle. Make sure aspirator tube is pushed
completely onto nipple of swivel & ball check assembly. |
| |
D) Replace
aspirator tube assembly. |
| |
E) Remove
solution bottle and aspirator tube from swivel & ball check
assembly. Gently insert small implement (paper clip, small
dia. hex key, etc.) into nipple of swivel & ball check assembly
to dislodge rubber ball check. |
|
|
|
| 5 |
PROBLEM: |
|
|
WATER LEAKING BACK INTO SOLUTION
BOTTLE |
|
|
|
CAUSE: |
|
|
A) Rubber
ball damaged or missing. |
|
|
B) Rubber
ball not sitting in seat of swivel & ball check assembly. |
|
|
|
SOLUTION: |
|
|
A) Replace
rubber ball using Deoseptic Diverter Rebuild Kit. Includes
Outlet Tube Gasket (item 6,) Spray Arm Repair Kit (item
11,) and all items in Pivot Replacement Kit (item 12)
except Pivot. |
|
|
B) Remove
swivel and ball check assembly, taking care not to lose rubber
ball. Insert rubber ball back into seat of swivel & ball
check assembly and reassemble. NOTE: Follow steps in solution
(E) of problem 4. |
|
| |
| |
| |
|