IIIIREPAIR SERVICEIIIIIIII PARTS & UNITS INDEXIII IIII ABOUT USIIIIIIII HOME
 
 

DOWNLOAD A PDF VERSION OF THIS GUIDE HERE.||||||

1A
# 54700
Pedestal Faucet Spout & Faucet& Sensor Assembly (EBF-615)
1B
# 54701
Faucet Spout & Sensor Assembly (EBF-650)
2
# 59173
Spray Head w/Key - (male 0.56 gpm)
# 54683
Aerator Spray Head w/Key (male2.2 gpm)
# 54902
Laminar Flow Spray Head (male 2.2 gpm )
3
# 54693
Key Only
4A
# 54703
Mounting Kit for EBF-615
4B
# 54678
Mounting Kit for EBF-650
5A
# 54685
1/8" NPT Pipe to 3/8" Compression
5B
# 54685
3/8" Tee Compression Fitting
5C
# 54680
Solenoid Supply Kit (NOT SOLD)
6
# 54325
Control Module Assembly
6A
# 54335
Solenoid Replacement Kit. Includes Solenoid Body, Solenoid Enclosure, and Solenoid Filter Cap
6B
# 8578
Solenoid Valve Module Assembly for Transformer Powered mode (-R variation.) Includes solenoid enclosure, solenoid body, and filter cap.
7
# 54686
Mounting Hardware
8
# 57970
Sensor Replacement Kit includes Cable Assembly, Retainer for PC Board Assembly, and Housing for Sensor Window
 
 
MAINTENANCE GUIDE FOR THE EBF-615, AND EBF-650 FAUCET
 
1
PROBLEM:
  Sensor troubleshooting LED does not function (red indicator light does not flash during set-up procedure.)
   
  CAUSES:
  A) No battery power is being supplied to sensor.
  B) Insufficient battery power is being supplied to sensor.
  C) Sensor Cable is not properly installed.
  D) Sense range is set at minimum distance.
  E) Control Module assembly is defective.
   
  SOLUTIONS:
  A) Ensure that the batteries are installed properly. Check that the orientation of each battery matches the positive (+) and negative (-) symbols shown in the bottom of the battery compartment. Reinsert the Battery Compartment control Module.
  B) One (or more) of the batteries is "dead". To ensure proper operation, insert four (4) new C-size alkaline batteries. Check that the orientation of each battery matches the positive (+) and negative (-) symbols shown in the bottom of the Battery Compartment. Reattach Battery Compartment to Control Module.
  C) Disconnect and reconnect Sensor Cable to the Control Module
  D) Increase Sensor range. Use a small phillips screwdriver to turn the potentiometer screw (white screw in a blue base) clockwise.
  E) Replace # 54325 (EBF-60-A) Control Module assembly.
   

2
PROBLEM:
  Faucet does not deliver any water when sensor is activated.
   
  INDICATOR:
  Solenoid valve produces audible "click".
   
  CAUSES:
  A) Water supply valve is closed.
   
  SOLUTION:
  A) Open the supply stop.
   
  INDICATOR:
  Solenoid Valve DOES NOT produce an audible "click".
   
  CAUSES:
  B) Solenoid lead is not properly connected to the Control Module.
  C) Batteries are not installed properly.
   
  SOLUTIONS:
  B) Disconnect and reconnect Solenoid Lead to the Control Module.
  C) Check that the orientation of each battery matches the positive (+) and negative (-) symbols shown in the bottom of the battery compartment. Reinsert the Battery Compartment control Module. The trouble shooting LED should flash RED when a user is detected.
3
PROBLEM:
  Faucet delivers only a slow flow or dribble when sensor is activated.
     
  CAUSE:  
  A) Water supply valve is partially closed.
  B) Solenoid Filter is clogged.
  C) Aerator or Spray Head is clogged.
     
  SOLUTION:  
  A) Completely open the Supply Stop(s) .
  B) Remove, clean, and reinsert. Replace with # 8578 (EBF-1004-A) if necessary
  C) Remove, clean, and reinsert
   

4
PROBLEM:
  Faucet does not stop delivering water or continues to drip after user is no longer detected (automatic shut-off fails even when batteries are removed.)
     
  CAUSE:  
  A) Solenoid Valve has been connected backwards.
  B) Solenoid Valve is dirty
  C) Solenoid Valve Module is defective.
     
  SOLUTIONS:
  A) Disassemble Solenoid Valve compression fittings at both the inlet and outlet positions. Water should flow from inlet through the Solenoid Valve to outlet according to the direction shown by the arrow on the side of the Solenoid Valve. Reconnect the fittings in the correct orientation.
  B) Backflush by reversing water flow (opposite to the direction shown by the arrow on the side of the Solenoid Valve) through the Solenoid Valve. Reconnect the compression fittings in the correct orientation. Activate faucet.
  C) Replace # 54335 (EBF-62-A) Solenoid Valve Module.
   
   

5
PROBLEM:
  The water temperature is too hot or too cold on a faucet connected to hot and cold supply lines with two Back Checks.
     
  CAUSES:  
  A) Supply Stops are not adjusted properly.
  B) One Back Check is installed backwards.
     
  SOLUTIONS:
  A) Adjust Supply Stops
  B) Reinstall Back Check
 
 
CARE AND CLEANING INSTRUCTIONS
 
DO NOT USE abrasive or chemical cleaners to clean faucets that may dull the luster and attack the chrome or special decorative finishes. Use ONLY soap and water, then wipe dry with a clean towel or cloth.
 
When cleaning the bathroom tile, protect the faucet from any splattering of cleaner. Acids and cleaning fluids can discolor or remove chrome plating.
 
 
 

DOWNLOAD A PDF VERSION OF THIS GUIDE HERE.|||

Manual Flushometers | Manual Flushometer Parts | Electronic Flushometers | Electronic Flushometer Parts | Electronic Faucets | Electronic Faucet Parts
 
 
© 2004 Equiparts Corp. All Rights Reserved.