SLOAN ELECTRONIC FAUCET REPAIR PARTS BREAKDOWN
EBF-615  •  EBF-650
EBF-615 and EBF-650 faucet repair parts
1A
# 54700
Pedestal Faucet Spout & Faucet& Sensor Assembly (EBF-615)
1B
# 54701
Faucet Spout & Sensor Assembly (EBF-650)
2
# 59173
Spray Head w/Key - (male 0.56 gpm)
# 54683
Aerator Spray Head w/Key (male2.2 gpm)
# 54902
Laminar Flow Spray Head (male 2.2 gpm )
3
# 54693
Key Only
4A
# 54703
Mounting Kit for EBF-615
4B
# 54678
Mounting Kit for EBF-650
5A
# 54685
1/8" NPT Pipe to 3/8" Compression
5B
# 54685
3/8"Tee Compression Fitting
5C
 
Solenoid Supply Kit (NOT SOLD)
6
# 54325
Control Module Assembly
6A
# 54335
Solenoid Replacement Kit. Includes Solenoid Body, Solenoid Enclosure, and Solenoid Filter Cap
6B
# 8578
Solenoid Valve Module Assembly for Transformer Powered mode (-R variation.) Includes solenoid enclosure, solenoid body, and filter cap.
7
# 54686
Mounting Hardware
8
# 57970
Sensor Replacement Kit includes Cable Assembly, Retainer for PC Board Assembly, and Housing for Sensor Window
 
MAINTENANCE GUIDE FOR THE EBF-615, AND EBF-650 FAUCET
1
PROBLEM:
Sensor troubleshooting LED does not function (red indicator light does not flash during set-up procedure.)
CAUSE:
A) No battery power is being supplied to Sensor
B) Insufficient battery power is being supplied to Sensor
C) Sensor Cable is not properly inserted
D) Sensor Range is set at minimum distance
E) Control Module assembly is defective
SOLUTIONS:
A) Ensure that the batteries are installed properly. Check that the orientation of each battery matches the positive and negative symbols shown on the bottom of the battery compartment. Reinsert the Battery Compartment into the Control Module.

B) One or more of the batteries is "dead." To ensure proper operation, replace with four C-size batteries
( order four # 32739 batteries from Equiparts by calling 800-442-6622.) Check that the orientation of each battery matches the positive and negative symbols shown on the bottom of the battery compartment. Reinsert the Battery Compartment into the Control Module.

C) Disconnect and reconnect Sensor Cable to the Control Module.
D) Increase Sensor range. Use a small phillips screwdriver to turn the potentiometer screw (white screw in a blue base) clockwise.
E) Replace EBF-60-A Control Module assembly by ordering part # 54325 from Equiparts. Call 800-442-6622 to order.
2
PROBLEM:
Faucet does not deliver any water when Sensor is activated.
INDICATOR:
A) Solenoid valve produces an audible "click."
CAUSE:
A) Water supply valve is closed
SOLUTION:
A) Open the supply stop.
INDICATOR:
B) Solenoid valve does NOT produce an audible "click."
CAUSE:
B) Solenoid lead is not properly connected to the Control Module
SOLUTION:
B) Disconnect and reconnect Solenoid lead to the Control Module
CAUSE:
C) Batteries are not installed properly.
SOLUTION:
C) Check that the orientation of each battery matches the positive and negative symbols shown on the bottom of the battery compartment. Reinsert the battery compartment into the control module. The troubleshooting LED should flash red when a user is detected
CAUSE:
D) Solenoid needs replaced
SOLUTION:
D) Order part # 54335 from Equiparts. To order, call 800-442-6622.
 
3
PROBLEM:
Faucet delivers only a slow flow or dribble when sensor is activated.
CAUSES:
A) Water supply valve is partially closed.
B) Solenoid Filter is clogged.
C) Aerator or Spray Head is clogged.
D) Aerator needs replaced.
SOLUTIONS:
A) Completely open the Supply Stop.
B) Remove, clean, and reinsert. Replace (EBF-1004-A) Solenoid Filter Kit if necessary by ordering part # 08578 from Equiparts. Call 800-442-6622 to order.
C) Remove, clean, and reinsert.
D) Call Equiparts at 800-442-6622 to order part # 59857.

4
PROBLEM:
Faucet does not stop delivering water or continues to drip after user is no longer detected (automatic shut-off fails even when batteries are removed.)
CAUSE:
A) Solenoid Valve has been connected backwards.
  B) Solenoid Valve is dirty.
  C) Solenoid Valve Module is defective.
 
SOLUTIONS:
A) Disassemble Solenoid Valve compression fittings at the inlet and outlet positions. Water should flow from inlet through the Solenoid Valve to outlet according to the direction shown by the arrow on the side of the Solenoid Valve. Reconnect the compression fittings in the correct orientation.
B) Back flush by reversing water flow opposite to the direction shown by the arrow on the side of the Solenoid Valve) through the Solenoid Valve. Reconnect the compression fittings in the correct orientation. Activate faucet.
C) Replace (EBF-62-A) Solenoid Valve Module by ordering part # 80754 from Equiparts. Call 800-442-6622 to order.

5
PROBLEM:
The water temperature is too hot or too cold on a faucet connected to hot and cold supply lines with two Back Checks.
CAUSES:
A) Supply Stops are not adjusted properly.
B) One Back Check is installed backwards.
SOLUTIONS:
A) Adjust Supply Stops
B) Reinstall Back Check
NOTE: For some systems, a thermostatic mixing valve may be required. Order part # 54621 from Equiparts. Call 800-442-6622 to order.
Equiparts  •   1003 Main St.  •  Pittsburgh, PA 15215
Pittsburgh: 412-781-9100   •  Toll Free: 800-442-6622   •   Fax: 412-781-3085
© 2008 Equiparts Corp. All Rights Reserved.